Description:
The main factors that cause the aging of materials is the sun and damp, xenon lamp test chamber use a xenon lamp simulating sunlight, condensation moisture simulating rain and dew, test the damage to material and product caused by sunlight, rain and dew.
The test materials were placed in a programmed recycling environment with certain temperature of light and moisture alternate, take a few days or weeks you can reproduce the harm of months or even years outdoor.
Artificial accelerated aging test data can help to choose new materials, transformation of existing materials, and evaluate how formula changes influence the durability of the product.
This equipment is wide used in many industries; say non-metal material, organic materials (paint, coating, and rubber, plastic), and new tech materials and so on.
Features:
1. Inner liner is made of high-grade stainless steel plate, A3 plate color powder.
2. Dynamic rotating sample holder, speed adjustable.
3. Touch screen programmable controller.
4. Double layer filtering observation window, harmful gas exhaust outlet.
5. Double metal blackboard thermometer.
6. Kinds of safety protection device.
Specifications:
Model | KP-XDW (water cooling) | KP-XDA (air cooling) |
Temperature | RT+10~63℃; experiment blackboard temperature: 65℃ ±3 ℃ | |
Rain time | 0~9999 min, adjustable | |
Rain duration | 0~240 min, Interval (break) adjustable | |
Wave-length | 290nm~800nm | |
Radiation intensity | 550W/m2 | |
Xenon lamp | 6KW, water cooling, life span 500H | 4.5 KW, air cooling, life span 2000H |
Volume | 950*950*850 mm | 800*800*800 mm |
Liner | SUS304 stainless steel | SUS304 stainless steel |
Shell | Spray paint cold rolling steel plate | |
Sample shelf | SUS304 stainless steel, rotate ≤5±2 RPM | Aluminum metal alloy, rotate ≤5±2 RPM |
Distance to the lamp | 480mm, effectively radiate 6500cm2 | 375mm, effectively radiate 6500cm2 |
Controller | Programmable touch screen | Brand Programmable touch screen |
Observation window | Double layer filtering observation window | |
Blackboard thermometer | Double metals blackboard thermometer | Double metals blackboard thermometer; Brand radiation meter |
Percolation slice | Domestic | International brand |
Power | 380V 50HZ | |
Standard | GB/T16422.4-1999 GB/T2423.24-1995 ASTM G 155-2005 | |