Formaldehyde and VOC Tester

Micro VOC Release Chamber

Product purpose       

• To detect substance released on the material surface in steady temperature.First,data of volatile substances and molecular materials before modification;Second,comparison of product properties(such as different colors or patterns)within the effective range;Third,detection of sample machine and low volatile materials;fourth,detection of the consensus of products during official authentication test.  

• Measure volatile and semivolatile organic compounds in major materials,and implement regular quality control.  

• Detect volume and radiations.        

• When temperature changes,detect the steam permeated into materials  

• Analysis of flavors and fragrances 

 

Function   

 

Micro sampling chamber can provide standard testing environment(temperature,cleanliness,relative humidity,air exchange rate,,etc) and detect VOC/SVOC and water content of various materials and products.It is very suitable for concrete analysis of heterogeneous prototype and detection of low trace elements.   

 

 

Reference Standard  

 

ISO 12219-3: 2012(E) air in road vehicles Section 3 Determination method for screening release emission of volatile organic components of materials of vehicle interiors with micro test chamber method.

 

Technical Parameters:

 

Product Model※1

MV-4

MV-6

MV-8

Environmental temperature

5~35℃

Environmental humidity

≤90%R.H

Requirements for Environmental

Cleanliness No pollutant source in the surrounding environment TVOC 500μg/m³        

Volume of climate chamber

Φ64*H34mm=116ml,inner arc surface chamber

Quantity of inner tanks※1

Serveral,with aspiration at the bottom ring and outlet and sampling at the ceiling

Adjustable range of temperature

60~200℃high temperature cleaning more than 260℃ 

Adjustable range of humidity(optional)

5%R.H(65℃)

Pressure

Positive pressure

Temperature fluctuation

±0.3℃(fluctuation)

Temperature deviation

±1.0℃(deviation)

Humidity fluctuation※2(optional)

±2.5%R.H

Humidity deviation※2(optional)

±3%

Air exchange rate

10~500/times±5%

Air velocity

10~500mL/min(quality flow controller

Power

Alternating current 220V/50Hz

Starting power※3

3kw

Operating power※3

1.5kw

External dimension of climate chamber※4

W580mm*D890*H1050(when there are 6 inner tanke),it is up to the quantity of inner tanks

Background concentration of methanol

smaller than 15μg/m³,the lowest 6μg/m³   

Background concentration of VOC

single VOC has a concentration smaller than 2 μg/m³,T VOC 20 μg m dust filtrated TVOC has a concentration smaller than 20μg/m³

Rate of recovery

Formaldehyde,octanol,4 phenyl cyclohexene,dodecane and tetradecane have a recovery rate larger than 80%

Clean air system

pressure>1.0Mpa,dry and clean gases(non combustible and noe toxic harmful gases such as air,nitrogen and argon)

Cooling mode

Air cooling

Controller

Color touch screen man-machine interface;Mitsubishi pld controller